Basic specifications
Accuracy guaranteed: 1 year
Max. number of connectable modules | 4 plug-in modules + 7 wireless modules | |||||
---|---|---|---|---|---|---|
Connectable modules (Plug-in modules) |
U8550, U8551, U8552, U8553, U8554, U8555 (U8555: supported V2.00 or later) |
|||||
Connectable modules (Wireless modules) |
LR8530, LR8531, LR8532, LR8533, LR8534, LR8535 (LR8535: supported V2.00 or later) |
|||||
No. of measurement channels | Up to 120 ch with plug-in modules, up to 330 ch with plug-in modules and wireless modules (U8555 and LR8535 can input up to 500 channels per module) |
|||||
Pulse/logic input | [Number of ch] 8 ch (common GND, non-isolated, exclusive setting for pulse/logic input for individual channels) [Adaptive input format] Non-voltage contact, open collector, or voltage input [Count] 0 to 1000 M pulse, 1 pulse resolution [Rotational speed] 0 to 5000/n (r/s), 1/n (r/s) resolution, 0 to 300,000/n (r/min.), 1/n (r/min.) resolution, n: Number of pulses per rotation (1 to 1000) [Logic input] Records 1 or 0 for each recording interval |
|||||
Recording intervals | 1 ms*, 2 ms*, 5 ms* (*: Can be set only when using 1 ms/S modules), 10 ms to 1 hour, 22 selections (Data refresh interval can be set for each module) | |||||
Data storage | SD Memory Card/USB Drive (user-selectable) (Only storage media sold by HIOKI are guaranteed for operation) |
|||||
LAN interface | 100BASE-TX / 1000BASE-T, DHCP, DNS support, Functions: Data acquisition, condition settings used with the Logger Utility software, configuring settings and controlling recording using communications commands, FTP server / FTP client, HTTP server, Email transmission, NTP client |
|||||
Wireless LAN interface | IEEE 802.11b/g/n Communications range: 30 m, line of sight Encryption function: WPA-PSK/WPA2-PSK, TKIP/AES Usable channels: 1 to 11 Supported modes: Wireless module connectivity, access point, station Functions: Configuring settings and controlling recording using communications commands, FTP server / client, HTTP server, Email transmission, NTP client |
|||||
USB interface | Series A receptacle × 2: USB 2.0 compliant (USB drive, keyboard, or hub)) Series mini-B receptacle × 1: Data acquisition, condition settings used with the Logger Utility, configuring settings and controlling recording using communications commands, transferring data from a connected SD Memory Card to a computer |
|||||
SD card slot | SD standard-compliant slot × 1 (with SD memory card/SDHC memory card support), Guaranteed-operation options: Z4001, Z4003 | |||||
Display | 7 inch TFT color liquid crystal display (WVGA 800 × 480 pixel) | |||||
Functions | Save waveform data in real time to the SD memory card or USB drive, numerical value calculations, waveform calculations, 8ch alarm output, voltage output ×2 (5 V /12 V /24 V selectable) | |||||
Power supply | [AC adapter] Using the Z1014 (100 to 240 V AC, 50/60 Hz), 95 VA Max. (including AC adapter), 28 VA Max. (exclusive of AC adapter) [Battery Pack] Using the Z1007 (accommodates 2 batteries), continuous use 4 hr (reference value for 2 pieces), 20 VA Max. [External power] 10 to 30 V DC, 28 VA Max. |
|||||
Dimensions and mass | Without any modules: 272 mm (10.71 in) W × 145 mm (5.71 in) H × 43 mm (1.69 in) D (excluding protrusions), 1108 g (39.1 oz) (excluding battery pack) With 2 modules: 272 mm (10.71 in) W × 198 mm (7.80 in) H × 63 mm (2.48 in) D (excluding protrusions) With 4 modules: 272 mm (10.71 in) W × 252 mm (9.92 in) H × 63 mm (2.48 in) D (excluding protrusions) |
|||||
Included accessories | Quick Start Manual ×1, LOGGER Application Disc (Quick Start Manual, Instruction Manual, Logger Utility, Logger Utility Instruction Manual, CAN editor, CAN editor instruction manual, Communication Instruction Manual) ×1, USB Cable ×1, AC Adapter Z1014 ×1, Precautions Concerning Use of Equipment that Emits Radio Waves ×1 |
Countries and regions where wireless operation is currently supported (LR8450-01, Z3230)
Last update: August 2023
Countries and regions where wireless operation is currently supported |
Australia, Austria, Bangladesh, Belgium, Brazil, Bulgaria, Canada, Chile, China, Colombia, Croatia, Cyprus, Czech Republic, Denmark, Egypt, Estonia, Finland, France, Georgia, Germany, Greece, Hong Kong, Hungary, Iceland, India, Indonesia, Ireland, Israel(*1), Italy, Japan, Kenya, Korea, Kuwait, Latvia, Liechtenstein, Lithuania, Luxembourg, Macao, Malaysia(*1), Malta, Myanmar, Netherlands, New Caledonia, New Zealand, Norway, Philippines, Poland, Portugal, Romania, Saudi Arabia, Singapore, Slovakia, Slovenia, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Kingdom, United Arab Emirates (UAE), United States, Viet Nam*1: The approval is valid for specific distributor or importer only. Please be consulted by our sales office for details. |
|||||
---|---|---|---|---|---|---|
Note |
Use of devices that emit radio radiation requires approval in the country of use. Thus, use of the instrument in a country or region where they have not been approved may be subject to penalty as a violation of law. |