KYOWA
UCAM-60

Features

The data logger UCAM-60C M14 is an all-in-one measuring instrument developed in full pursuit of easier field measurement.
It has easy-to-operate keys, a bright readable display providing understandable presentation and a printer for immediate confirmation of measurement results

Description

Summary

・Saving the long-term saving measured results to the built-in memory (Approx. 1.8 GB)
・USB memory ensures easy data collection
・Up to 1000 channels measurement
(With external scanners USB-70B series)
・Measurement up to 20000 ×10-6 strain with a resolution of 0.1 ×10-6 strain (With full bridge system)
・All-in-one measuring instrument (UCAM-60C M14)
・Online data logger fully controlled from the PC (UCAM-65C M14)

Specifications

Measuring Targets Strain gages
Strain-gage transducers
DC voltage-output instruments
DC current-output instruments
Civil engineering transducers with a thermal sensor
Potentiometer sensors
Thermal sensors (Thermocouples and platinum resistance thermometer bulbs)
Connectable Scanners USS-61B, USS-62B, USS-63B
(Dedicated scanners, mounted on top of the UCAM-60C M14)The main unit is connected to the following scanners via the optional scanner interface (USI-67A).
USB-70B series (via USI-67A)
Channels Max. 30 with dedicated scanners
Max. 1000 with external scanners connected
Max. 1000 with dedicated scanners and external scanners connected
Input Terminals Can connect to lead wires through either soldering or screwing.
(Screw: M3 with metal plate)NDIS4102 (7 pins) connectors (USS-62B)
Switching Terminal Semiconductor relays
Scanning Speed 50 ms/channel (Standard mode)
0.28 s/channel (High-resolution mode)
20 ms/channel (High-speed mode)——-Scanners——————–Line Frequencies————
Dedicated scanner (Standard mode)
50 Hz Zone: 50 ms/channel
60 Hz Zone: 50 ms/channel
Dedicated scanner (High-resolution mode)
50 Hz Zone: 0.28 s/channel
60 Hz Zone: 0.28 s/channel
Dedicated scanner (High-speed mode)
50 Hz Zone: 20 ms/channel
60 Hz Zone: 20 ms/channel
USB-70B (Standard mode only)
50 Hz Zone: 60 ms/channel
60 Hz Zone: 58.4 ms/channel[Note 1]
Scanning speeds stated above are standard maximum speeds in respective modes.
Besides these, the following speeds are set for each individual channel:
0.28 s, 0.5 s, 1 s, 2 s, 5 s, and 10 s.[Note 2]
High-resolution mode and high-speed mode are selectable for dedicated scanners only.[Note 3]
High-resolution mode is individually switchable for desired channels.
High-speed mode is only collective switching for all channels of dedicated scanners.
Repeat measurement interval time
= (Number of Measuring channels × scanning speed)
+ data processing time (2 to 20 s)
*Data processing time is indeterminate, changed by measurement setting and environment.——-Measuring Targets——————–Scanning Speed————
Strain (Gage & transducer)
Standard Mode (50 ms/channel)
High-resolution Mode (0.28 s/channel)
High-speed Mode (20 ms/channel)
Voltage/current-output sensor
Standard Mode (50 ms/channel)
High-speed Mode (20 ms/channel)
Civil engineering transducer
Standard Mode (50 ms/channel)
Temperature sensor (TC, Pt)
Standard Mode (50 ms/channel)
Potentiometer sensor
Standard Mode (50 ms/channel)
High-speed Mode (20 ms/channel)[Note 1]
High-resolution mode is available only with full bridge system.
[Note 2]
High-speed mode is available full bridge system, voltage, current, and potentiometer sensor.
Operating Modes Real-time
Monitor
Automatic (Interval measurement)
Measurement Functions Initial (Initial values are measured and stored in internal memory.)
Measure (Initial values are subtracted from original values.)
Original (Raw values are measured without subtraction of initial values.)
Easy Measure (Auto zero balancing function is activated.)*The selected function is applied to all channels.
Coefficient Calculation Function Multiplication by calibration coefficient
Calibration by TEDS
Conversion of measured values to physical quantities
Scaling
Correction
Unit 59 units
Automatic Measurement Function
Interval Measurement Measurement is automatically performed at preset time intervals.
Trigger Measurement A relative value (certain changing quantity) or an absolute value triggers measurement.
Trigger Interval Measurement Combination of trigger measurement and interval measurement.
Storage Internal memory
Capacity: Approx. 1.8 GB
Strain Measurement (Standard Mode)
Constant Voltage Excitation Approx. 2 or 5 VDC
Constant Current Excitation Approx. 5.7 mA (Bridge resistance 350 Ω)
Approx. 16.7 mA (Bridge resistance 120 Ω)
Scanning Speed 50 ms/channel
Gage Factor 2.00 fixed
(Coefficient calculation function enables correction with 2.00/Ks.)
Initial Value Memory Range Same as measuring range.
Measuring Range, Resolution and Accuracy Measuring Range: 0 to ±50 k ×10-6 strain
Resolution: 1 ×10-6 strain
Accuracy: ±(0.05% of reading +1) ×10-6 strainMeasuring Range: ±50 k to ±500 k ×10-6 strain
Resolution: 10 ×10-6 strain
Accuracy: ±(0.05% of reading +10) ×10-6 strain*Resolution and accuracy be automatically changed by Autorange function.
Strain Measurement (High-resolution Mode)
Constant Voltage Excitation Approx. 5 VDC
Constant Current Excitation Approx. 16.7 mA (Bridge resistance 350 Ω)
Scanning Speed 0.28 s/channel
Initial Value Memory Range Same as measuring range.
Gage Factor 2.00 fixed
(Coefficient calculation function enables correction with 2.00/Ks.)
Measuring Range, Resolution and Accuracy Measuring Range: 0 to ±20 k ×10-6 strain
Resolution: 0.1 ×10-6 strain
Accuracy: ±(0.05% of reading +0.3) ×10-6 strainMeasuring Range: ±20 k to ±200 k ×10-6 strain
Resolution: 1 ×10-6 strain
Accuracy: ±(0.05% of reading +3) ×10-6 strainNote 1: Available only with full bridges system.
Note 2: Bridge resistance should be 120 to 1000 Ω for bridge excitation with constant voltage.
Note 3: Bridge resistance should be 350 Ω for bridge excitation with constant current.
[Measuring range]
0 to ±15000 ×10-6 strain
0 to ±150000 ×10-6 strain
Note 4: Available only with dedicated scanners.
Note 5: Resolution and accuracy be automatically changed by Autorange function.
Strain Measurement (High-speed Mode)
Constant Voltage Excitation Approx. 2 VDC
Constant Current Excitation Approx. 5.7mA (Bridge resistance 350 Ω)
Approx. 16.7mA (Bridge resistance 120 Ω)
Scanning Speed 20 ms/channel
Gage Factor 2.00 fixed
(Coefficient calculation function enables correction with 2.00/Ks.)
Initial Value Memory Range Same as measuring range.
Measuring Range, Resolution and Accuracy Measuring Range: 0 to ±50 k ×10-6 strain
Resolution: 1 ×10-6 strain
Accuracy: ±(0.08% of reading +3) ×10-6 strainMeasuring Range: ±50 k to ±500 k ×10-6 strain
Resolution: 10 ×10-6 strain
Accuracy: ±(0.08% of reading +30) ×10-6 strain[Note 1]
Available only with full bridges system (120 to 1000 Ω).
[Note 2]
Available only with dedicated scanners.
[Note 3]
Resolution and accuracy be automatically changed by Autorange function.
Voltage Measurement (Standard Mode)
Scanning Speed 50 ms/channel
Initial Value Memory Range Same as measuring range
Measuring Range, Resolution and Accuracy Channel Mode: V/500mV
Measuring Range: 0 to ±50.000 mV
Resolution: 1 μV
Accuracy: ±(0.05% of reading +0.003) mV
Input Resistance: 10 MΩ or moreChannel Mode: V/500mV
Measuring Range: ±50.00 to ±500.00 mV
Resolution: 10 μV
Accuracy: ±(0.05% of reading +0.03) mV
Input Resistance: 10 MΩ or moreChannel Mode: V/50V
Measuring Range: 0 to ±5.0000 V
Resolution: 100 μV
Accuracy: ±(0.05% of reading +0.0002) V
Input Resistance: 1 MΩ or moreChannel Mode: V/50V
Measuring Range: ±5.000 to ±50.000 V
Resolution: 1 mV
Accuracy: ±(0.05% of reading +0.002) V
Input Resistance: 1 MΩ or more*Resolution and accuracy be automatically changed by Autorange function.
Voltage Measurement (High-speed Mode)
Scanning Speed 20 ms/channel
Initial Value Memory Range Same as measuring range
Measuring Range, Resolution and Accuracy Channel Mode: V/500mV
Measuring Range: 0 to ±50.000 mV
Resolution: 1 μV
Accuracy: ±(0.08% of reading +0.006) mV
Input Resistance: 10 MΩ or moreChannel Mode: V/500mV
Measuring Range: ±50.00 to ±500.00 mV
Resolution: 10 μV
Accuracy: ±(0.08% of reading +0.06) mV
Input Resistance: 10 MΩ or moreChannel Mode: V/50V
Measuring Range: 0 to ±5.0000 V
Resolution: 100 μV
Accuracy: ±(0.08% of reading +0.0006) V
Input Resistance: 1 MΩ or moreChannel Mode: V/50V
Measuring Range: ±5.000 to ±50.000 V
Resolution: 1 mV
Accuracy: ±(0.08% of reading +0.006) V
Input Resistance: 1 MΩ or more[Note 1]
Resolution and accuracy be automatically changed by Autorange function.
[Note 2]
Available only with dedicated scanners.
Current Measurement (Standard Mode)
Scanning Speed 50 ms/channel
Initial Value Memory Range Same as measuring range
Measuring Range, Resolution and Accuracy Channel Mode: I/50 mA
Measuring Range: 0 to ±50.00 mA
Resolution: 10 μA
Accuracy: ±(0.05% of reading +0.01) mA[Note 1]
External shunt resistor (high-accuracy 250 Ω) is required.
[Note 2]
Stated accuracy does not include the resistance error of the external shunt resistor.
Current Measurement (High-speed Mode)
Scanning Speed 20 ms/channel
Initial Value Memory Range Same as measuring range
Measuring Range, Resolution and Accuracy Channel Mode: I/50 mA
Measuring Range: 0 to ±50.00 mA
Resolution: 10 μA
Accuracy: ±(0.08% of reading +0.01) mA[Note 1]
Available only with dedicated scanners.
[Note 2]
External shunt resistor (high-accuracy 250 Ω) is required.
[Note 3]
Stated accuracy does not include the resistance error of the external shunt resistor.
Temperature Measurement with Thermocouple (Standard Mode)
Scanning Speed 50 ms/channel
Measuring Range, Resolution and Accuracy 【Type: K】
Measuring Range: −200.0 to 1230.0℃
Resolution: 0.1℃
Accuracy: ±0.7℃
Internal Reference Junction Compensator Accuracy: ±0.5℃
(With input terminal temperature balanced in an ambient)
(Temp. range of 0 to 50°C)【Type: T】
Measuring Range: −200.0 to 400.0℃
Resolution: 0.1℃
Accuracy: ±0.7℃
Internal Reference Junction Compensator Accuracy: ±0.5℃
(With input terminal temperature balanced in an ambient)
(Temp. range of 0 to 50°C)【Type: E】
Measuring Range: −200.0 to 660.0℃
Resolution: 0.1℃
Accuracy: ±0.5℃
Internal Reference Junction Compensator Accuracy: ±0.5℃
(With input terminal temperature balanced in an ambient)
(Temp. range of 0 to 50°C)【Type: J】
Measuring Range: −200.0 to 870.0℃
Resolution: 0.1℃
Accuracy: ±0.6℃
Internal Reference Junction Compensator Accuracy: ±0.5℃
(With input terminal temperature balanced in an ambient)
(Temp. range of 0 to 50°C)【Type: R】
Measuring Range: 0 to 1760.0℃
Resolution: 0.1℃
Accuracy: ±2.2℃
Internal Reference Junction Compensator Accuracy: ±0.5℃
(With input terminal temperature balanced in an ambient)
(Temp. range of 0 to 50°C)[Note 1]
Accuracies do not include the internal reference junction compensator accuracy.
[Note 2]
The reference junction compensator is switchable between internal and external.
[Note 3]
Thermocouple resistance should be 1 kΩ or less.
Temperature Measurement with Civil Engineering Transducers with a Thermal Sensor (Standard Mode)
Scanning Speed 50 ms/channel
Measuring Range, Resolution and Accuracy Measuring Range: −50.0 to 200.0℃
Resolution: 0.1℃
Accuracy: ±0.5℃[Note 1]
Target physical quantity and temperature are measured in a single channel.
[Note 2]
Strain measuring range are the same as in strain measurement in standard mode.
Temperature Measurement with Platinum Resistance Thermometer Bulb (Standard Mode)
Scanning Speed 50 ms/channel
Measuring Range, Resolution and Accuracy 【Type: Pt100】
Measuring Range: −200.0 to 660.0℃
Resolution: 0.1℃
Accuracy: ±0.3℃【Type: JPt100】
Measuring Range: −200.0 to 510.0℃
Resolution: 0.1℃
Accuracy: ±0.3℃Note: Connection is 3-wire system.
Measurement with Potentiometer Sensor (Standard Mode)
Scanning Speed 50 ms/channel
Initial Value Memory Range Same as measuring range
Sensor Power Supply Approx. 2 VDC
Potentiometer Resistance 1 kΩ to 10 kΩ
Measuring Range, Resolution and Accuracy Channel Mode: POT.
Measuring Range: 0 to ±50.00%
Resolution: 0.01%
Accuracy: ±0.1% FS
Measurement with Potentiometer Sensor (High-speed Mode)
Scanning Speed 20 ms/channel
Initial Value Memory Range Same as measuring range
Sensor Power Supply Approx. 2 VDC
Potentiometer Resistance 1 kΩ to 10 kΩ
Measuring Range, Resolution and Accuracy Channel Mode: POT.
Measuring Range: 0 to ±50.00%
Resolution: 0.01%
Accuracy: ±0.1% FSNote: Available only with dedicated scanners.
Internal Timer Real time clock is built-in. (Battery backup)
Display Fluorescent display tube
128 × 64 dots
Printer Printing: Thermal
Paper width: 58 mm (24 characters/line)
Printing speed: 60 mm/s (Max.)
Interface RS-232C
LAN (10BASE-T/100BASE-TX)
USB2.0 (Collects measurement data by USB Flash Drive.)*1: Measurement data cannot be saved directly to USB memory.
*2: USB Flash Drives
Capacity: 32 GB or less
File Format: FAT32
*3: Recommend USB Flash Drives: GH-UFI-XSD2G
(Manufacturer: GREENHOUSE)
File Conversion Binary measurement data can be converted to CSV.
Self Diagnosis Functions Checks display, printer, bridge excitation, leadwire-off, input/output resistance, insulation resistance, mode, etc.
TEDS [Interface]
IEEE1451.4 Mixed Mode Transducer Interface Class2[Applicable sensor]
Should have information written in accordance with IEEE template No.33.
Cable length should be 30m or less.*With dedicated scanner USS-61B/62B/63B.
Operating Temperature 0 to 50ºC
Operating Humidity 20% to 85% (Non-condensing)
Storage Temperature -20 to 60ºC
Setting Maintenance Function ACOM at measurement circuit is switchable between floating and GND connect.
Power Supply 100 to 240 VAC (AC-operated version)

10 to 16 VDC (DC-operated version)
*DC operated version has power control function.

Current Consumption 100 VAC: 0.5 A or less (With 3 dedicated scanners mounted)
12 VDC: 4.0 A or less (With 3 dedicated scanners mounted)
Dimensions 360 W ×88 H ×400 D mm (Excluding protrusions)
Weight Approx. 6.3 kg (Excluding scanner)
Approx. 9.6 kg (With 3 dedicated scanners USS-62B mounted)
Standard Accessories AC power cable P-18 (With 2-pin conversion plug CM-52)
(AC-operated version)
DC power cable P-76
(DC-operated version)
Recording paper UCAM-60A-RP (1 roll)
Screwdriver
Spare fuse
CD-R (Instruction Manual)
Ground wire P-72
Optional Accessories Recording paper UCAM-60A-RP (10 rolls/pack)

Dimensions

UCAM-60C