KYOWA
WGI-470AS1
Instrumentation Amplifier

  • High-speed sampling: 2000 times/s
  • 2-analog output: Simultaneous voltage and current output, individual scaling
  • Peak hold: 5 types of peak hold
  • 3 sensitivity registrations: actual load calibration, sensitivity registration calibration, numerical value registration calibration

Description

WGI-470AS1, Instrumentation Amplifier

The WGI-470AS1 is a compact and high-speed sampling Instrumentation Amplifier which has optimized functions required for measurement in combination with strain-gage transducers.

Models:

  • WGI-470AS1-00: No option
  • WGI-470AS1-01: BCD output
  • WGI-470AS1-02: RS-232C
  • WGI-470AS1-03: RS-485

Channels: 1

Applicable Transducers: Strain-gage transducers

Compatible Bridge Resistance: 87.5 to 1000 Ω (350 Ω, 4 in parallel connection)

Bridge Excitation: 2 VDC, 10 VDC, switchable

Measuring Range: ±3.2 mV/V (Including zero adjustment range)

Zero Adjustment Range: Within measuring range (Not retained when power supply interrupted)

Sampling Speed: 2000 times/s

Resolution: 64000 counts/input range in both polarities

Calibration Functions:

  • Actual load calibration
  • Sensitivity registration calibration
  • Numerical value registration calibration

Display Functions:

  • Main display: 5 digits (±99999), polar LED
  • Sub display: 1 digit (9)
  • Update speed: 0.13, 0.25, 0.51, 1.02, 2.03, 4.07, 8.16, and 16.28 s (Thinning display)
  • Comparative determination: 3 points (HIGH, OK, LOW)
  • Status display: 2 points (HOLD, LOCK)

Nonlinearity: Within ±(0.02%FS ±1 digit)

Smoothing Functions:

  • Minimum scale : 1, 2, 5, 10, 20, 50, and 100 counts
  • Moving average : 0, 2, 4, 8, 16, 32, 64, 128, 256, 512, 1024, and 2024 times
  • Filters : 10, 30, 100 Hz and FLAT

Comparison/Judgment Functions:

  • Comparators : 2 points (HIGH, LOW)
  • Hysteresis : Individual width settings possible
  • Comparison mode : Normal/at hold
  • Judgment : 1 point

Pattern Functions: Registers 4 groups of pattern files (comparative values) and enables switching through setting of functions.

Control Input: 4 points:

  • Zero command
  • Level test command
  • Hold command
  • Reset command

Control Output: 3 points:

  • HIGH
  • OK
  • LOW