Description
WGA-120A, Carrier Type Instrumentation Amplifier
The WGA-120A is a Carrier Type Instrumentation Amplifier, which is suitable for connecting a strain-gage transducer to measure loads, pressure, torques or displacements
Models:
- WGA-120A-00
- Power Codes: DC
- Power Supply: 10.5 to 15 VDC
- Power Consumption: 3.5 W or less
- Balance Adjustment: Manual
- WGA-120A-01
- Power Codes: 100 VAC
- Power Supply: 90 to 110 VAC
- Power Consumption: 6.5 VA or less
- Balance Adjustment: Manual
- WGA-120A-02
- Power Codes: 200 VAC
- Power Supply: 180 to 220 VAC,
- Power Consumption: 6.5 VA or less
- Balance Adjustment: Manual
- WGA-120A-03
- Power Codes: 240 VAC
- Power Supply: 216 to 264 VAC
- Power Consumption: 6.5 VA or less
- Balance Adjustment: Manual
- WGA-120A-10
- Power Codes: DC
- Power Supply: 10.5 to 15 VDC
- Power Consumption: 3.5 W or less
- Balance Adjustment: Automatic
- WGA-120A-11
- Power Codes: 100 VAC
- Power Supply: 90 to 110 VAC
- Power Consumption: 6.5 VA or less
- Balance Adjustment: Automatic
- WGA-120A-12
- Power Codes: 200 VAC
- Power Supply: 180 to 220 VAC
- Power Consumption: 6.5 VA or less
- Balance Adjustment: Automatic
- WGA-120A-13
- Power Codes: 240 VAC
- Power Supply: 216 to 264 VAC
- Power Consumption: 6.5 VA or less
- Balance Adjustment: Automatic
Specifications
Channels: 1
Applicable Transducers: Strain-gage transducers
Compatible Bridge Resistance:
- 87.5 to 1000 Ω (Up to four 350 Ω transducers connected in parallel.
- Bridge excitation is limited to 2 or 1 Vrms for transducers with bridge resistance less than 175 Ω.)
Bridge Excitation: 5, 2, 1 Vrms, square wave (Switchable by changing internal jumper connection)
Input Range
[Sensitivity Switch ×2000]
- Bridge Excitation(Vrms):1
- Input Range (mV/V)Sensitivity Adjustment×1 to ×0.4:±5 to ±12.5
- Bridge Excitation(Vrms):2
- Input Range (mV/V)Sensitivity Adjustment×1 to ×0.4:±2.5 to ±6.25
- Bridge Excitation(Vrms):5
- Input Range (mV/V)Sensitivity Adjustment×1 to ×0.4:±1 to ±2.5
[Sensitivity Switch ×4000]
- Bridge Excitation(Vrms):1
- Input Range (mV/V)Sensitivity Adjustment×1 to ×0.4:±2.5 to ±6.25
- Bridge Excitation(Vrms):2
- Input Range (mV/V)Sensitivity Adjustment×1 to ×0.4:±0.5 to ±1.25
- Bridge Excitation(Vrms):5
- Input Range (mV/V)Sensitivity Adjustment×1 to ×0.4:±0.5 to ±1.25
[Sensitivity Switch ×10000]
- Bridge Excitation(Vrms):1
- Input Range (mV/V)Sensitivity Adjustment×1 to ×0.4:±1 to ±2.5
- Bridge Excitation(Vrms):2
- Input Range (mV/V)Sensitivity Adjustment×1 to ×0.4:±2.5 to ±6.25
- Bridge Excitation(Vrms):5
- Input Range (mV/V)Sensitivity Adjustment×1 to ×0.4:±0.2 to ±0.5
Rated Output
- Voltage: ±10 V (Load resistance: 2 k Ω or more)
- Current: 4 to 20 mA (Load resistance: 500 Ω or less, voltage output: 0 to 10 V)
- (Switchable by changing internal jumper connection)
Zero Adjustment Range: Within ±1.5 mV/V, manual balance or auto balance by selecting model
Sensitivity Adjustment Range: ×2000, ×4000, ×10000 (Switchable by changing internal jumper connection) Adjustable between ×0.4 and ×1.0 by the trimmer.
Frequency Response: 10, 30, 100, or 500 Hz (Switchable by changing internal jumper connection)
SN Ratio: 53dB p-p or more
Stability:
Temperature:
Zero point: Within ±0.2μVRTI /°C
Sensitivity: Within ±0.05%/°C
Time
Zero point: Within ±0.5μVRTI /8h
Sensitivity: Within ±0.2%/8h
Conditions
Sensitive: ×10000, ×1.0 (Adjustment by the trimmer)
Bridge resistance: 120 Ω
Bridge excitation: 2 Vrms
EDP-70, DIN Rail Mounting Fixture