Description
WD-100A, Indicator for Voltage-Output Type Sensors
Displays voltage/current and pulse on single unit
Measurement Channel: 2
Display:
- 2.4 inch TFT LCD
- Ach measurement result, Bch measurement result, calculation result,
- Ach and Bch measurement results,
- Ach or Bch measurement result and calculation result
Over Warning: OVER or -OVER when display range are exceeded
External Controls: Following 5 functions can be assigned to control terminals (user-configurable).
- Comparator reset function
- Totalized value reset function
- Measurement prohibited function: Measurement prohibited A/B/A&B
- Current value hold function: Current value hold A/B/A&B
- Max value hold function: Max value hold A/B/A&B
- Min value hold function: Min value hold A/B/A&B
- Digital zero function
- Pattern change function: Pattern change 1 to 3
- Display change function
- Trend hold function: As follows, only shortcut setting
- Compare list function
Input Specifications
Analog Input Measurement of Ach
Measurement Range:
- Measurement range: ±5 V
- Input impedance: About 1 MΩ
- Maximum allowable input: ±100 V
- Accuracy: ±(0.05% of FS + 1 digit)
- Measurement range: 0 to 5 V
- Input impedance: About 1 MΩ
- Maximum allowable input: ±100 V
- Accuracy: ±(0.05% of FS + 1 digit)
- Measurement range: 1 to 5 V
- Input impedance: About 1 MΩ
- Maximum allowable input: ±100 V
- Accuracy: ±(0.05% of FS + 1 digit)
- Measurement range: ±10 V
- Input impedance: About 1 MΩ
- Maximum allowable input: ±100 V
- Accuracy: ±(0.05% of FS + 1 digit)
- Measurement range: 0 to 10 V
- Input impedance: About 1 MΩ
- Maximum allowable input: ±100 V
- Accuracy: ±(0.05% of FS + 1 digit)
- Measurement range: 4 to 20 mA
- Input impedance: About 10 Ω
- Maximum allowable input: ±50 mA
- Accuracy: ±(0.05% of FS + 1 digit)
- Measurement range: 0 to 20 mA
- Input impedance: About 10 Ω
- Maximum allowable input: ±50 mA
- Accuracy: ±(0.05% of FS + 1 digit)
- Measurement range: ±20 mA
- Input impedance: About 10 Ω
- Maximum allowable input: ±50 mA
- Accuracy: ±(0.05% of FS + 1 digit)
*Each range can measure up to ±10% FS range.
(Internal limit processing with ± 10% FS.)
The full scale in the bipolar input setting considers plus and minus separately. For example, in the case of ± 10V input, limit processing is performed up to ±11 V. (20 V is not treated as FS.)
Similarly, the accuracy with ±10 V input is also specified as one-sided FS treatment, and the accuracy is calculated as 5 mV (0.05%) ± 1 digit.
Pulse Input Measurement of Bch
Frequency Range: 0.01 Hz to 250 kHz
Input Signal: Open collector (NPN/PNP), voltage pulse, totem pole output (complementary output), AC pulse, proximity sensor
Input Method: Single-phase pulse
SamplingRate: 10 ms (calculation period)
Display Updating Period: 100 ms